28-Aug-90 11:00 AM CST
Performance of Retrieved Kuntscher Intramedullary Rods: Improved Corrosion Resistance with Contemporary Material Design
Clinical Materials, Vol. 5, pp. 53-71, 1990
Author: Stephen D. Cook, Mark R. Brinker, Ronald C. Anderson, Robert J. Tomlinson & James Co Butler
Click here to read the entire published article by Dr Brinker
Eighteen Kuntscher-type intramedullary rods were obtained after human implantation periods ranging from 1 to 23 years. A clinical evaluation, metallurgical and chemical analyses of each implant were performed. Idiopathic, implant-related pain was the most common reason for removal, but was not related to extent of implant corrosion. When classified as early versus contemporary material compositions, a significant increase in surface corrosion, inclusion content and carbon content were demonstrated in rods made of the earlier composition of stainless steel. Intergranular corrosion was associated with structural design, rather than material microcleanliness or composition. While contemporary intramedullary fracture fixation provides excellent clinical results, the present investigation suggests that the performance of stainless steel implants may be improved through continued refinement of steel composition, careful consideration of structural design and fabrication techniques, and by routine removal of implants at the earliest opportunity.
Click a star to rate!
Rating: 0.00 / 5.00
-
0 ratings
Related Documents:
Performance of Retrieved Kuntscher Intramedullary Rods.pdf 39.551 KB
Content Tags: bone •
Other Recent Articles:
- Case Report Calcaneogenesis 1-May-09
- Management of Falls After Total Knee Arthroplasty 1-Mar-08
- Ilizarov Distraction Before Revision Hip Arthroplasty After Resection Arthroplasty With Profound Limb Shortening 1-Jan-08
- Outcomes of Tibial Nonunion in Older Adults Following Treatment Using the Ilizarov Method 1-Oct-07
- Metabolic and Endocrine Abnormalities in Patients With Nonunions 1-Sep-07
- Ilizarov Treatment of Infected Nonunions of the Distal Humerus After Failure of Internal Fixation: An Outcomes Study 1-Mar-07
- Payer Type Has Little Effect on Operative Rate and Surgeons’ Work Intensity 1-Oct-06
- Nonunions of the Femoral Shaft and Distal Femur 11-Aug-06
- CT Scan Helps Delineate Cervical Osteoid Osteoma and Atypical Nidus 17-Sep-05
- IIizarov Correction of Malrotated Femoral Shaft Fracture Initially Treated With an Intramedullary Nail: A Case Report 17-Sep-04
Return to the Dr. Mark Brinker, Houston Orthopedic Surgeon Articles Search Page